Strona 1 - CL1000 Ceiling Light

OWNER’S MANUALCL1000 Ceiling LightMarus11727 Fruehauf DriveCharlotte, NC 28273 USA

Strona 2 - Warranty

812”12”18”18”MOUNTING LOCATIONFOR RIGHT-HANDEDOPERATORMOUNTING LOCATIONFOR LEFT-HANDEDOPERATORTYPICAL POSITION OF CEILING MOUNTIN RELATION TO DENTAL C

Strona 3 - Table of Contents

91 INCHCEILING JOISTS16” O.C.POWER SUPPLYCONDUITFINISHED CEILINGMATERIALMOUNTING PLATEw/ COLLARPOWER CONDUIT COMES THROUGH HOLE 2-1/2” OFFSET FROM CEN

Strona 4 - General Information

10CEILING JOISTFIXED CEILINGSUSPENDEDCEILINGSUSPENDEDCEILINGCOVERCOVERMOUNTING PLATEw/ COLLAR & POSTMOUNTING PLATEw/ COLLAR & POSTSUGGESTED IN

Strona 5 - Safety Notes

11Metal Structure for Suspended CeilingElectrical RequirementsThe 115VAC/230VAC service line for the ceiling light should be evaluated and installed b

Strona 6 - Technical Information

12Ceiling Light Installation Instructions Although traditional methods dictate mounting the light to the left side of the chair for a right-han

Strona 7 - Cleaning the Light

1311. Check the light for freedom of movement and for any drift when positioned. Recheck the post and make any necessary leveling adjustments. Raise t

Strona 8 - CLEANING YOUR LIGHT

14Power SupplyAccess HoleJunction BoxPost SocketSuspension Tube orCeiling Post(Refer to chart on Pg. 4)GroundTerminalSafety BoltCeiling MountingPlateC

Strona 9 - Dimensions

15Overview / Replacement PartsControlsThe on-off switch is located on the control panel directly above the light reector. The long toggle provides

Strona 10 - Suggested Mounting Methods

16 Turn 1/8 Clockwise to lock - logo should be level TetherAdjustmentsBulb Removal and ReplacementNote: For maximum reliability and optimum operation

Strona 11 - EXISTING CONSTRUCTION

17Light Focus AdjustmentsNote: Focus adjustment should be checked each time a new bulb is installed.1. Place a white paper or cardboard sheet in fron

Strona 12 - FIGURE 6

Marus11727 Fruehauf DriveCharlotte, NC 28273 Printed in the U.S.A. 70-050R228 Rev 11 02/12Technical SupportTech Support: 800-304-5332FAX

Strona 13 - Electrical Requirements

18Yoke Pivot Friction AdjustmentHorizontalNote: If the light head pivots too freely horizontally, an adjustment can be made as follows:1. Remove the

Strona 14

19Flex Arm AdjustmentVertical1. Remove the two ex arm cover screws using a at head screwdriver. (See gure 15).2. Remove the ex arm cover.3. Usi

Strona 15 - Figure 8

20ELECTROMAGNETIC COMPATIBILITYElectrical medical devices are subject to special EMC safety measurements and as a result the equipment must be install

Strona 16 - Figure 9

21This product is intended for use in the electromagnetic environment specied below. The customer or the user of this product should ensure that it

Strona 17 - Controls

22NOTE 1: At 80 MHz to 800 MHz, the higher frequency range applies.NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagati

Strona 18 - Adjustments

23Guidance and manufacturer‘s declaration-electromagnetic emissionsThis product is suitable for use in all establishments, other than domestic establi

Strona 19 - Light Focus Adjustments

24Electrical SchematicsCL1000 Ceiling Light - 115V 0V115V230V

Strona 20 - Horizontal

25Electrical SchematicsCL1000 Ceiling Light- 230V 230V115V0V

Strona 21 - Flex Arm Adjustment

26Verify the following after installation or servicing of the light: All manuals are present. All labels are present and legible. No mechanical dam

Strona 23

1Table of ContentsGeneral Information ...

Strona 25

29Purchase InformationNotes / Service HistoryWrite in the model and serial numbers below for all applicable equipment such as the chair, unit light an

Strona 26 - Electrical Schematics

70-050R228Rev. 11 •02/12Marus11727 Fruehauf DriveCharlotte, NC 28273 USATechnical support: 800-304-5332Web Site: www.marus.com

Strona 27

2The following symbols may be used throughout this manual:CAUTION. Failure to carefully follow the described procedure may result in damage to the eq

Strona 28 - CHECKLIST

3Storage Conditions:-55°C to +50°C10% to 90% Relative HumidityThe pre-installation must be performed according to the requirements in our ‘Pre-install

Strona 29

4Technical InformationTechnical Specications for the Dental LightTechnical Description — LightThe dental light is used for illuminating the oral cav-

Strona 30

5WARNINGS, DISINFECTING & STERILIZATION - LIGHTSInfection control in the dental ofce continues to be a high priority for our customers and end us

Strona 31 - Purchase Information

6CLEANING YOUR LIGHTCleaning Reector and CoverNOTE: Use only denatured alcohol to clean the glass reector. Use only mild soap and water to clean th

Strona 32 - Web Site: www.marus.com

712”23 1/2”21-1/2” - 26”28”TRAVELDimensionsLocationA typical location for mounting the ceiling mount light is relation to the dental chair for a right